5.8.2 Procedure
The organization shall maintain a documented procedure for the control of TMMDE. The procedure shall include requirements for the specific equipment type and shall address:
a) unique identification;
b) calibration status;
c) traceability to international or national measurement standards; where no such standards exist, the basis used for calibration shall be recorded;
d) calibration method and acceptance criteria;
e) frequency of calibration, and when the calibration interval begins;
f) documentation of the calibration measurements prior to adjustment and measurements after any adjustments during calibration;
NOTE Calibration measurements prior to adjustment can be referred to as ‘as-found’. Calibration measurements after any adjustments can be referred to as ‘as-left’. When no adjustments are made, ‘as-found’ and ‘as-left’ are the same.
g) actions taken to prevent unintended use of TMMDE identified as out-of-calibration, beyond calibration interval, or not in-service;
h) when the TMMDE is found to be out of calibration, an assessment of the validity of previous measurements and actions to be taken on the TMMDE and product, including maintaining records and evidence of notification to the customer if suspect product has been shipped;
i) use of third-party, proprietary, employee-owned, and customer-owned TMMDE;
j) maintenance; and
k) suitability for the planned monitoring and measurement activities.
Controlling testing, measuring, monitoring, and detection equipment is critical for several reasons in the oil and gas industry:
- Accuracy and Reliability: Ensuring that measurements and test results are accurate and reliable is essential for maintaining product quality, safety, and regulatory compliance.
- Compliance: Adherence to industry standards and regulatory requirements often mandates strict control over such equipment to avoid legal and financial penalties.
- Safety: Accurate monitoring and detection equipment are vital for ensuring safe operations, preventing accidents, and protecting personnel and the environment.
- Operational Efficiency: Well-maintained and calibrated equipment helps in achieving operational efficiency, minimizing downtime, and reducing maintenance costs.
- Quality Assurance: Controlling equipment ensures that products meet specified requirements and customer expectations consistently.
Steps to Control Testing, Measuring, Monitoring, and Detection Equipment
- Identification:
- Assign unique identification numbers or codes to each piece of equipment.
- Maintain an inventory list with detailed information about each item, including manufacturer, model, serial number, and location.
- Calibration:
- Establish and document calibration intervals based on manufacturer recommendations, industry standards, historical performance data, and risk assessments.
- Use calibration standards traceable to national or international standards.
- Record calibration dates, results, and any adjustments made.
- Maintenance:
- Implement a preventive maintenance program specifying the type, frequency, and responsibilities for maintaining equipment.
- Keep detailed maintenance records, including dates, types of maintenance performed, and results.
- Verification and Validation:
- Perform periodic verification checks to ensure equipment remains within specified tolerances.
- Validate new or repaired equipment before putting it back into service to ensure it meets all required standards.
- Handling and Storage:
- Develop procedures for the proper handling of equipment to prevent damage.
- Store equipment in designated areas with appropriate environmental controls to protect it from conditions that could affect its accuracy and longevity.
- Out-of-Tolerance Equipment:
- Clearly identify and label equipment that is found to be out of tolerance.
- Remove out-of-tolerance equipment from service and take corrective actions, such as recalibration or repair.
- Assess the impact on previous measurements and document all actions taken.
- Training:
- Provide training to personnel on the proper use, calibration, and maintenance of equipment.
- Maintain records of training and competence assessments.
- Record-Keeping:
- Maintain comprehensive records of all calibration, maintenance, verification, and validation activities.
- Ensure records are easily accessible and regularly updated.
- Regular Audits:
- Conduct regular audits of the control processes to ensure compliance with internal procedures and external standards.
- Use audit findings to continuously improve the equipment control processes.
Equipment Required by Oil and Gas Industries
In the oil and gas industry, a wide range of equipment is used for testing, measuring, monitoring, and detection. Some common types include:
- Testing Equipment:
- Pressure Testers
- Flow Testers
- Leak Detectors
- Measuring Equipment:
- Pressure Gauges
- Flow Meters
- Temperature Sensors
- Level Meters
- Dimensional Calipers
- Monitoring Equipment:
- Condition Monitoring Systems (vibration, temperature, pressure)
- Environmental Monitors (air quality, water quality)
- Data Loggers
- Detection Equipment:
- Gas Detectors (methane, hydrogen sulfide, carbon monoxide)
- Smoke Detectors
- Flame Detectors
- Ultrasonic Leak Detectors
Controlling testing, measuring, monitoring, and detection equipment is crucial for ensuring the accuracy, reliability, safety, and compliance of operations in the oil and gas industry. By implementing systematic steps such as identification, calibration, maintenance, verification, proper handling, and thorough record-keeping, organizations can maintain high standards of quality and safety. Additionally, a variety of specialized equipment is required to support these activities, tailored to the specific needs of the oil and gas sector.
The organization must maintain a documented procedure for the control of TMMDE.
Maintaining a documented procedure for the control of testing, measuring, monitoring, and detection equipment is crucial for ensuring compliance, consistency, quality, efficiency, and risk management. By developing comprehensive procedures that cover all aspects of equipment control, from identification to verification and record-keeping, an organization can ensure its operations are reliable, safe, and efficient. This not only helps in meeting regulatory and industry standards but also supports the overall quality and performance of the organization’s products and services. The Organization must maintain a documented procedure for the control of testing, measuring, monitoring, and detection equipment for the following reasons:
- Compliance with Standards and Regulations:
- API Q1 and Other Standards: Maintaining a documented procedure is often a requirement of industry standards such as API Q1, ISO 9001, and other regulatory frameworks. This ensures that the organization adheres to best practices and meets legal obligations.
- Regulatory Compliance: Ensures compliance with governmental and industry regulations, thereby avoiding legal penalties and ensuring the safety and reliability of operations.
- Consistency and Reliability:
- Standardized Processes: Documented procedures provide a standardized approach to equipment control, ensuring that all personnel follow the same methods, which enhances consistency and reliability of measurements.
- Minimized Errors: Reduces the likelihood of human error by providing clear instructions and guidelines for equipment use, calibration, maintenance, and verification.
- Quality Assurance:
- Product Quality: Ensures that products meet the required specifications and quality standards by maintaining accurate and reliable testing and measurement equipment.
- Traceability: Provides traceability of measurements and calibrations, which is essential for quality assurance and auditing purposes.
- Operational Efficiency:
- Preventive Maintenance: Scheduled maintenance and calibration prevent unexpected equipment failures, reducing downtime and associated costs.
- Effective Resource Management: Efficient use of resources by ensuring that equipment is only out of service for calibration and maintenance when necessary.
- Risk Management:
- Safety: Ensures the safety of operations by verifying that monitoring and detection equipment function correctly, preventing accidents and hazardous situations.
- Risk Mitigation: Identifies potential issues before they lead to significant problems, thereby mitigating risks to the organization and its stakeholders.
How to Maintain a Documented Procedure for the Control of Testing, Measuring, Monitoring, and Detection Equipment
- Developing the Procedure:
- Scope and Purpose: Clearly define the scope and purpose of the procedure, specifying the types of equipment covered and the goals of the control process.
- Responsibilities: Assign responsibilities for equipment control, including calibration, maintenance, and record-keeping, to specific roles within the organization.
- Identification and Documentation:
- Equipment Identification: Establish a system for uniquely identifying each piece of equipment, including its location, manufacturer, model, and serial number.
- Documentation System: Implement a documentation system (manual or electronic) to record all relevant information about each piece of equipment, including calibration schedules, maintenance records, and verification results.
- Calibration and Maintenance:
- Calibration Schedule: Define and document calibration intervals based on manufacturer recommendations, industry standards, and historical data.
- Maintenance Plan: Develop and document a preventive maintenance plan that outlines the type, frequency, and procedures for maintaining each piece of equipment.
- Verification and Validation:
- Verification Checks: Document procedures for periodic verification checks to ensure ongoing accuracy and reliability of equipment.
- Validation Processes: Include validation processes for new or repaired equipment to confirm they meet all required specifications before use.
- Handling and Storage:
- Handling Procedures: Document procedures for the proper handling of equipment to prevent damage and ensure accuracy.
- Storage Conditions: Specify and document appropriate storage conditions to protect equipment from environmental factors that could affect performance.
- Out-of-Tolerance Equipment:
- Identification and Labeling: Clearly document how out-of-tolerance equipment will be identified and labeled.
- Corrective Actions: Outline the steps to be taken when equipment is found out of tolerance, including recalibration, repair, and impact assessment.
- Training and Competence:
- Training Programs: Document training programs to ensure personnel are competent in using, calibrating, and maintaining equipment.
- Competence Records: Maintain records of training and competence assessments for all relevant personnel.
- Record-Keeping:
- Detailed Records: Document the requirement for maintaining detailed records of all calibration, maintenance, verification, and validation activities.
- Audit Trail: Ensure the documentation system provides a clear audit trail for internal and external audits.
- Review and Improvement:
- Periodic Reviews: Schedule and document periodic reviews of the equipment control process to ensure it remains effective and up-to-date.
- Continuous Improvement: Document procedures for incorporating feedback and audit findings into process improvements.
Example of a Procedure for Control of Equipment
1. Purpose: To ensure all testing, measuring, monitoring, and detection equipment is controlled, calibrated, and maintained to produce accurate and reliable measurements.
2. Scope: This procedure applies to all equipment used in the inspection, testing, monitoring, and detection processes within the organization.
3. Responsibilities:
- Quality Manager: Overall responsibility for equipment control and compliance with API Q1 requirements.
- Calibration Technician: Responsible for performing calibrations, maintenance, and keeping records.
- Department Managers: Ensure equipment within their departments is properly used, maintained, and calibrated.
4. Procedure:
4.1 Identification and Traceability:
- Assign a unique identification number to each piece of equipment.
- Record the identification number, manufacturer, model, serial number, date of acquisition, and location in the equipment database.
4.2 Calibration:
- Calibrate equipment at intervals defined by the calibration schedule.
- Use calibration standards traceable to national or international standards.
- Record calibration results, including any adjustments made.
4.3 Maintenance:
- Perform preventive maintenance according to the maintenance schedule.
- Document maintenance activities, including the date, type of maintenance, and results.
4.4 Verification and Validation:
- Conduct periodic verification checks to ensure ongoing accuracy.
- Validate new or repaired equipment before use in production.
4.5 Handling, Storage, and Preservation:
- Handle equipment with care to prevent damage.
- Store equipment in designated areas with controlled environmental conditions.
- Implement preservation measures for equipment not in use.
4.6 Out-of-Tolerance Equipment:
- Clearly label equipment that is out of tolerance and remove it from service.
- Investigate and document the impact of out-of-tolerance equipment on previous measurements.
- Take corrective actions, including recalibration or repair.
4.7 Training:
- Provide training for personnel on proper use, calibration, and maintenance of equipment.
- Maintain records of training and competence assessments.
5. Records: Maintain records of all calibrations, maintenance activities, verification checks, out-of-tolerance investigations, and training.
6. Review and Audit: Conduct regular reviews and audits of equipment control processes to ensure compliance and identify areas for improvement.
The procedure must include requirements for the specific equipment type.
To include requirements for specific equipment types in the control of testing, measuring, monitoring, and detection equipment, an organization should develop a detailed and structured approach. This involves several steps to ensure that all relevant factors are considered and that procedures are tailored to meet the needs of different equipment categories.
- Equipment Identification and Classification:
- Inventory Management: Create and maintain an inventory of all equipment, including details such as type, manufacturer, model, serial number, and location.
- Categorization: Classify equipment based on their type (testing, measuring, monitoring, detection), criticality, and application. This helps in tailoring specific requirements for each category.
- Requirement Definition:
- Regulatory and Standards Compliance: Identify relevant standards (e.g., API Q1, ISO 9001) and regulatory requirements that apply to each equipment type.
- Manufacturer Guidelines: Incorporate manufacturer recommendations for calibration, maintenance, and use.
- Risk Assessment: Conduct risk assessments to determine the criticality of each equipment type and its impact on product quality and safety.
- Developing Detailed Procedures:
- Calibration: Define specific calibration requirements for each equipment type, including:
- Frequency: Based on usage, manufacturer guidelines, and risk assessment.
- Standards: Use of traceable standards for calibration.
- Documentation: Recording calibration results, adjustments, and calibration certificates.
- Maintenance: Establish maintenance procedures specific to each equipment type, including:
- Preventive Maintenance: Scheduled activities such as cleaning, part replacement, and lubrication.
- Corrective Maintenance: Procedures for addressing and documenting repairs and adjustments.
- Verification and Validation: Set up verification and validation processes for ensuring ongoing accuracy and reliability, including:
- Periodic Checks: Regular verification intervals.
- Initial Validation: Validation of new or repaired equipment before use.
- Handling and Storage: Develop handling and storage guidelines to prevent damage and degradation, specific to the needs of each equipment type.
- Calibration: Define specific calibration requirements for each equipment type, including:
- Documentation and Record Keeping:
- Procedure Manuals: Develop and distribute detailed procedure manuals for each equipment type, outlining the specific requirements and steps to be followed.
- Record Templates: Create standardized templates for recording calibration, maintenance, verification, and validation activities.
- Audit Trails: Ensure that all documentation provides a clear audit trail for internal and external audits.
- Training and Competence:
- Training Programs: Develop and implement training programs tailored to the specific requirements of each equipment type.
- Competence Assessment: Regularly assess the competence of personnel in handling, calibrating, and maintaining different types of equipment.
- Refresher Courses: Offer periodic refresher courses to keep personnel updated on best practices and procedural changes.
- Monitoring and Review:
- Performance Monitoring: Continuously monitor the performance and condition of equipment to ensure compliance with specific requirements.
- Internal Audits: Conduct regular internal audits to verify adherence to procedures and identify areas for improvement.
- Feedback Mechanism: Establish a feedback mechanism for operators and maintenance personnel to report issues and suggest improvements.
- Continuous Improvement:
- Data Analysis: Analyze data from calibration, maintenance, and verification records to identify trends and areas for improvement.
- Procedure Updates: Regularly update procedures based on audit findings, feedback, and technological advancements.
- Best Practices: Incorporate industry best practices and lessons learned into procedures and training programs.
Example Framework for Specific Equipment Types
| Equipment Type | Calibration Requirements | Maintenance Requirements | Verification Requirements | Handling & Storage | Training Requirements |
|---|---|---|---|---|---|
| Pressure Gauges | – Calibrate every 6 months – Traceable standards – Document results | – Clean monthly – Annual inspection – Record maintenance | – Monthly verification – Validate new gauges | – Store in protective cases – Avoid impact | – Use and handling – Calibration procedures |
| Flow Meters | – Calibrate quarterly – Manufacturer standards – Document adjustments | – Monthly cleaning – Bi-annual inspection – Record maintenance | – Quarterly checks – Initial validation | – Store upright – Protect from dust and moisture | – Installation and use – Calibration methods |
| Gas Detectors | – Calibrate before each use – Follow manufacturer guidelines – Record calibration | – Weekly functional checks – Annual service – Record activities | – Pre-use verification – Validate after repair | – Store in controlled environment – Avoid exposure to contaminants | – Safety protocols – Maintenance and calibration |
By following these structured steps and using a framework like the one outlined above, an organization can effectively include and manage the specific requirements for different types of testing, measuring, monitoring, and detection equipment. This ensures that all equipment operates correctly, reliably, and in compliance with industry standards and regulatory requirements.
The procedure must include unique identification.
Including unique identification in the control of testing, measuring, monitoring, and detection equipment is crucial for maintaining accurate records, ensuring traceability, and managing calibration and maintenance schedules. Here’s a structured approach to achieving this:
Steps to Include Unique Identification
- Assign Unique Identification Numbers:
- ID System: Develop a system for assigning unique identification numbers to each piece of equipment. This system could include a combination of letters and numbers to denote the type of equipment, the location, and a serial number (e.g., PG-001 for Pressure Gauge number 1).
- Database Registration: Record these unique IDs in a centralized database, along with detailed information about each piece of equipment.
- Labeling and Tagging:
- Permanent Labels: Attach durable, permanent labels or tags to each piece of equipment. These labels should be resistant to environmental conditions such as moisture, heat, and chemicals.
- Barcodes/QR Codes: Include barcodes or QR codes on the labels to facilitate easy scanning and quick access to the equipment’s history and status in the database.
- Inventory Management:
- Database Integration: Integrate the unique IDs into an inventory management system that tracks the equipment’s location, status, calibration dates, maintenance records, and usage history.
- Regular Updates: Regularly update the database to reflect any changes in equipment status, location, or ownership.
- Tracking and Monitoring:
- Electronic Tracking: Use electronic tracking systems to monitor the location and status of each piece of equipment in real-time.
- Movement Logs: Maintain logs of equipment movements between locations or departments, ensuring that the unique ID is recorded in each log entry.
- Documentation and Record Keeping:
- Detailed Records: Keep detailed records for each piece of equipment, including purchase date, manufacturer, model number, unique ID, calibration history, maintenance records, and any incidents or repairs.
- Accessible Records: Ensure that these records are easily accessible to relevant personnel for reference and audits.
- Training and Awareness:
- Personnel Training: Train staff on the importance of unique identification, how to read and use the IDs, and how to update the system when changes occur.
- Standard Operating Procedures (SOPs): Develop and implement SOPs that detail the processes for assigning, labeling, and maintaining unique IDs.
- Audits and Verification:
- Internal Audits: Conduct regular internal audits to verify that all equipment is correctly identified and that records are up-to-date.
- Verification Processes: Implement verification processes to ensure the accuracy and consistency of unique IDs across all records and physical equipment.
Example of Unique Identification Records
| Equipment Type | Unique ID | Manufacturer | Model Number | Location | Last Calibration Date | Next Calibration Due |
|---|---|---|---|---|---|---|
| Pressure Gauge | PG-001 | ABC Instruments | PG-X100 | Lab 1 | 2024-01-10 | 2025-01-10 |
| Flow Meter | FM-002 | XYZ FlowTech | FM-2000 | Lab 2 | 2023-12-15 | 2024-06-15 |
| Gas Detector | GD-003 | GasSafe Inc. | GD-300 | Storage Room | 2024-05-01 | 2024-05-31 |
| Temperature Sensor | TS-004 | TempCo | TS-4000 | Lab 3 | 2023-11-01 | 2024-11-01 |
| Multimeter | MM-005 | MultiMeasure | MM-500 | Workshop | 2024-03-20 | 2024-09-20 |
Implementation Tips
- Consistent Application: Ensure the unique identification system is consistently applied to all relevant equipment.
- Technology Utilization: Use technology (such as RFID tags or QR codes) to streamline tracking and information retrieval.
- Regular Reviews: Periodically review and update the unique identification system to accommodate new equipment and retire old or obsolete items.
- Cross-Department Coordination: Coordinate between departments to ensure that all relevant equipment is included in the unique identification system and that records are harmonized.
The procedure must include calibration status.
Including calibration status in the control of testing, measuring, monitoring, and detection equipment is essential for ensuring the accuracy and reliability of measurements and tests. Here’s a structured approach to achieve this:
Steps to Include Calibration Status
- Identification and Labeling:
- Calibration Tags/Labels: Attach calibration tags or labels to each piece of equipment indicating its calibration status. The label should include the last calibration date, the next due date, and the unique identification number of the equipment.
- Color Coding: Use color-coded labels to quickly identify the calibration status (e.g., green for calibrated, yellow for nearing due date, red for overdue).
- Documentation and Records:
- Calibration Certificates: Maintain calibration certificates for each piece of equipment, detailing the calibration results, date, and the person who performed the calibration.
- Calibration Logs: Keep a log of all calibrations, including dates, results, any adjustments made, and the next due date for calibration. This log can be maintained electronically or in physical form.
- Automated Systems and Software:
- Calibration Management Software: Implement software to manage calibration schedules, send reminders for upcoming calibrations, and store calibration records. This software can provide real-time status updates and generate reports.
- Barcoding and Scanning: Use barcodes or QR codes on equipment labels to link directly to calibration records in the management system, making it easy to access and update information.
- Schedule and Planning:
- Calibration Schedule: Develop a calibration schedule based on the manufacturer’s recommendations, industry standards, and the criticality of the equipment. The schedule should be detailed and include all equipment that requires calibration.
- Frequency Determination: Determine calibration frequency based on usage patterns, equipment stability, past performance, and risk assessment.
- Training and Awareness:
- Personnel Training: Train personnel on the importance of calibration, how to read calibration labels, and how to update calibration status. This ensures that everyone is aware of the calibration status and follows the proper procedures.
- Awareness Programs: Conduct regular awareness programs and refresher courses to emphasize the significance of maintaining accurate calibration status.
- Monitoring and Verification:
- Periodic Checks: Conduct periodic checks to verify that the calibration status is up-to-date and that labels are correctly attached. This can be part of regular internal audits.
- Audit Trails: Ensure that there is an audit trail for each piece of equipment, documenting all calibrations, adjustments, and status changes.
- Non-Conforming Equipment:
- Identification and Isolation: Clearly identify and isolate any equipment that is out of calibration or overdue for calibration. This equipment should be labeled as “Out of Service” and should not be used until recalibrated.
- Corrective Actions: Implement corrective actions for any non-conforming equipment, including recalibration and investigation of potential impacts on past measurements.
Example of Including Calibration Status
| Equipment | ID Number | Last Calibration Date | Next Calibration Due | Calibration Status | Comments |
|---|---|---|---|---|---|
| Pressure Gauge | PG-101 | 2024-01-10 | 2025-01-10 | Calibrated | – |
| Flow Meter | FM-202 | 2023-12-15 | 2024-06-15 | Calibrated | – |
| Gas Detector | GD-303 | 2024-05-01 | 2024-05-31 | Nearing Due | Calibration due soon |
| Temperature Sensor | TS-404 | 2023-11-01 | 2024-11-01 | Calibrated | – |
| Multimeter | MM-505 | 2024-03-20 | 2024-09-20 | Calibrated | – |
Implementation Tips
- Consistency: Ensure consistent labeling and record-keeping practices across all equipment.
- Automation: Leverage automation tools to manage calibration schedules and records efficiently.
- Accessibility: Make calibration records easily accessible to relevant personnel for quick reference and verification.
- Regular Audits: Conduct regular audits to ensure compliance with calibration procedures and to identify any discrepancies.
The procedure must include traceability to international or national measurement standards, where no such standards exist, the basis used for calibration shall be recorded.
To ensure traceability to international or national measurement standards in the control of testing, measuring, monitoring, and detection equipment, an organization must establish and document a systematic approach. Here’s how to include this traceability, and handle situations where no such standards exist:
Steps to Ensure Traceability
- Identify Relevant Standards:
- International/National Standards: Identify applicable international (e.g., ISO, IEC) or national (e.g., NIST in the USA, NPL in the UK) measurement standards for each type of equipment.
- Industry Standards: If international or national standards are not available, refer to industry-specific standards or guidelines.
- Accredited Calibration Services:
- Accredited Laboratories: Use calibration services from laboratories accredited by recognized accreditation bodies (e.g., ISO/IEC 17025 accredited labs) to ensure that calibrations are traceable to national or international standards.
- Certification and Documentation: Obtain and maintain calibration certificates from these laboratories that explicitly state the traceability to recognized standards.
- Documentation of Traceability:
- Calibration Certificates: Ensure that calibration certificates include references to the standards used for calibration.
- Traceability Records: Maintain detailed records showing how each calibration is traceable to national or international standards. This should include the calibration certificate, the date, the standard used, and the laboratory details.
- Calibration Procedures:
- Standard Operating Procedures (SOPs): Develop SOPs that outline the process of calibration, including references to the standards used.
- In-House Calibration: If calibration is performed in-house, ensure that the procedures and equipment used are traceable to recognized standards, and document this traceability.
- Handling Non-Standard Situations:
- Alternative References: When no national or international standards exist, identify and document alternative references or methods used for calibration.
- Documentation of Basis: Record the basis used for calibration, including the method, any internal standards developed, and validation data supporting the accuracy and reliability of these methods.
- Regular Review and Verification:
- Periodic Review: Regularly review and verify calibration records and traceability documentation to ensure ongoing compliance.
- Audit and Validation: Conduct internal audits and validation checks to confirm that all calibrations are correctly documented and traceable.
Example of Traceability Documentation
| Equipment | ID Number | Calibration Date | Next Calibration Due | Standard Used | Calibration Certificate | Notes |
|---|---|---|---|---|---|---|
| Pressure Gauge | PG-101 | 2024-01-10 | 2025-01-10 | NIST Standard 1234 | Cert# NIST-1234-2024 | Calibrated by ABC Labs |
| Flow Meter | FM-202 | 2023-12-15 | 2024-06-15 | ISO 5167-1:2003 | Cert# ISO-5167-2023 | Calibrated by XYZ FlowTech |
| Gas Detector | GD-303 | 2024-05-01 | 2024-05-31 | Internal Method (No Standard) | Cert# INT-GD-2024 | Based on manufacturer’s guidelines |
| Temperature Sensor | TS-404 | 2023-11-01 | 2024-11-01 | ASTM E230 | Cert# ASTM-230-2023 | Calibrated by TempCo |
| Multimeter | MM-505 | 2024-03-20 | 2024-09-20 | IEC 61010-1 | Cert# IEC-61010-2024 | Calibrated by MultiMeasure |
Implementation Tips
- Comprehensive Records: Ensure that records are comprehensive and include all relevant details such as the calibration standard, method, date, and certificate number.
- Clear Documentation: Clearly document any deviations or use of alternative methods, providing justification and validation for these approaches.
- Regular Training: Train personnel on the importance of traceability and how to maintain accurate records.
- Quality Management System Integration: Integrate traceability documentation into the organization’s quality management system to ensure consistency and compliance.
The procedure must include calibration method and acceptance criteria.
To include the calibration method and acceptance criteria in the control of testing, measuring, monitoring, and detection equipment, an organization should follow a structured approach that integrates these elements into their standard operating procedures (SOPs) and quality management system. Here’s a detailed guide on how to achieve this:
Steps to Include Calibration Method and Acceptance Criteria
- Develop Standard Operating Procedures (SOPs):
- Calibration Method Documentation: Clearly document the calibration method for each type of equipment. This should include step-by-step instructions on how the calibration is performed, reference standards used, and any specific techniques or tools required.
- Acceptance Criteria Definition: Define the acceptance criteria for each piece of equipment. This should include the permissible tolerance levels, measurement uncertainty, and any specific performance indicators that the equipment must meet to be considered calibrated.
- Integration into Quality Management System:
- Quality Manual Inclusion: Incorporate the calibration methods and acceptance criteria into the organization’s quality manual or relevant quality management documents.
- Procedure References: Ensure that the SOPs for calibration reference the quality manual and vice versa, creating a cohesive system of documentation.
- Calibration Records:
- Detailed Records: Maintain detailed calibration records that include the calibration method used and whether the equipment met the acceptance criteria. These records should capture the date of calibration, the person who performed the calibration, the equipment details, and the results.
- Traceability: Ensure that each calibration record is traceable to the specific equipment through unique identification numbers.
- Training and Competency:
- Staff Training: Train personnel on the specific calibration methods and acceptance criteria for the equipment they handle. Ensure they understand how to document the calibration process and results accurately.
- Competency Verification: Regularly verify the competency of personnel conducting calibrations through assessments and audits.
- Regular Review and Updates:
- Periodic Review: Regularly review the calibration methods and acceptance criteria to ensure they remain current with industry standards and best practices.
- Continuous Improvement: Update the SOPs and quality manual as needed based on feedback, audit findings, and advancements in calibration techniques.
Example of Calibration Method and Acceptance Criteria Documentation
Calibration Method for Pressure Gauge
- Objective:
- To calibrate the pressure gauge and ensure its accuracy within specified tolerances.
- Equipment Required:
- Reference pressure standard (traceable to NIST)
- Pressure pump
- Data recording device
- Procedure:
- Connect the pressure gauge to the pressure pump.
- Incrementally increase the pressure and record the gauge reading at each step (e.g., 0 psi, 10 psi, 20 psi, etc.).
- Compare the readings of the pressure gauge with the reference standard.
- Record the deviations at each pressure level.
- Acceptance Criteria:
- The pressure gauge readings must be within ±1% of the reference standard across the entire pressure range.
- Any reading outside this tolerance indicates a need for adjustment or repair.
Calibration Method for Temperature Sensor
- Objective:
- To calibrate the temperature sensor and ensure its accuracy within specified tolerances.
- Equipment Required:
- Reference temperature standard (traceable to NIST)
- Temperature-controlled bath
- Data recording device
- Procedure:
- Immerse the temperature sensor and the reference standard in the temperature-controlled bath.
- Stabilize the bath at specific temperatures (e.g., 0°C, 25°C, 50°C, etc.).
- Record the readings of the temperature sensor and the reference standard.
- Compare the readings and calculate the deviations.
- Acceptance Criteria:
- The temperature sensor readings must be within ±0.5°C of the reference standard at all tested temperatures.
- Deviations outside this range require recalibration or sensor replacement.
Example of Calibration Record in Tabular Form
| Equipment Type | ID Number | Calibration Date | Calibrated By | Method Used | Acceptance Criteria | Result | Next Calibration Due |
|---|---|---|---|---|---|---|---|
| Pressure Gauge | PG-101 | 2024-01-10 | John Smith | Incremental Pressure Comparison | ±1% of reference standard | Within Tolerance | 2025-01-10 |
| Temperature Sensor | TS-202 | 2023-12-15 | Jane Doe | Temperature-Controlled Bath | ±0.5°C of reference standard | Within Tolerance | 2024-12-15 |
| Multimeter | MM-303 | 2024-05-01 | John Smith | Electrical Measurement | ±2% of reference standard | Outside Tolerance (Recalibrated) | 2024-11-01 |
| Flow Meter | FM-404 | 2023-11-01 | Jane Doe | Flow Rate Comparison | ±2% of reference standard | Within Tolerance | 2024-11-01 |
| Gas Detector | GD-505 | 2024-03-20 | John Smith | Gas Concentration Comparison | ±0.5% of reference standard | Within Tolerance | 2024-09-20 |
Implementation Tips
- Consistency: Ensure that calibration methods and acceptance criteria are consistently applied across all relevant equipment.
- Clear Communication: Communicate the importance of adherence to calibration methods and acceptance criteria to all relevant personnel.
- Periodic Audits: Conduct regular audits to ensure compliance with documented calibration procedures and acceptance criteria.
- Record Accuracy: Maintain accurate and detailed calibration records that are easily accessible for review and audits.
The procedure must include frequency of calibration, and when the calibration interval begins.
To include the frequency of calibration and the start date of the calibration interval in the control of testing, measuring, monitoring, and detection equipment, an organization should integrate these details into their documented procedures and quality management system. Here’s how to effectively manage these aspects:
Steps to Include Frequency of Calibration
- Determine Calibration Frequency:
- Manufacturer Recommendations: Base the initial calibration frequency on the equipment manufacturer’s guidelines.
- Industry Standards: Follow industry standards and best practices for specific types of equipment.
- Usage History: Adjust the frequency based on the equipment’s usage history and past performance.
- Risk Assessment: Perform a risk assessment to determine the impact of equipment failure on product quality and safety. Higher risk may necessitate more frequent calibration.
- Document the Calibration Frequency:
- Calibration Schedule: Develop a calibration schedule that specifies the frequency for each piece of equipment.
- Quality Manual: Include the calibration frequency in the organization’s quality manual or relevant SOPs.
- Calibration Records: Ensure that calibration records explicitly state the calibration frequency.
Steps to Determine When Calibration Interval Begins
- Date of First Use:
- Initial Calibration Date: Document the date of the initial calibration.
- First Use Date: If the calibration interval is based on the date of first use, record this date accurately.
- Subsequent Calibration Dates:
- Scheduled Intervals: Determine subsequent calibration dates based on the initial calibration date or the first use date and the specified calibration interval.
- Event-Based Triggers: In some cases, calibration intervals may be adjusted based on specific events, such as equipment repair or relocation. Document these events and their impact on the calibration schedule.
- Tracking and Notifications:
- Calibration Management System: Use a calibration management system to track calibration due dates and send notifications for upcoming calibrations.
- Log Entries: Maintain a calibration log that records each calibration event, including the date, the technician who performed the calibration, and the next due date.
Implementation in the Quality Management System
- Standard Operating Procedures (SOPs):
- Detailed Procedures: Develop SOPs that detail the process for determining and documenting calibration frequency and intervals.
- Templates and Forms: Use standardized templates and forms to capture calibration dates and intervals consistently.
- Training and Competency:
- Staff Training: Train personnel on how to determine calibration frequency and document calibration intervals accurately.
- Competency Checks: Perform regular competency checks to ensure that personnel follow the documented procedures correctly.
- Regular Review and Adjustment:
- Periodic Review: Regularly review calibration frequencies and intervals to ensure they remain appropriate based on equipment performance and changes in usage or risk.
- Adjustments: Adjust calibration schedules as needed based on review findings and ensure that all changes are documented and communicated.
Example of Calibration Schedule Documentation
| Equipment Type | ID Number | Calibration Frequency | Initial Calibration Date | Date of First Use | Next Calibration Due | Calibrated By | Notes |
|---|---|---|---|---|---|---|---|
| Pressure Gauge | PG-101 | Annually | 2024-01-10 | 2024-01-15 | 2025-01-10 | John Smith | N/A |
| Temperature Sensor | TS-202 | Semi-Annually | 2023-12-15 | 2024-01-05 | 2024-06-15 | Jane Doe | Adjusted frequency based on risk |
| Multimeter | MM-303 | Every 6 months | 2024-05-01 | 2024-05-03 | 2024-11-01 | John Smith | New equipment |
| Flow Meter | FM-404 | Annually | 2023-11-01 | 2023-11-05 | 2024-11-01 | Jane Doe | N/A |
| Gas Detector | GD-505 | Quarterly | 2024-03-20 | 2024-03-25 | 2024-06-20 | John Smith | High-risk equipment |
Key Points for Implementation
- Clear Documentation: Ensure that all relevant dates and frequencies are clearly documented in calibration records.
- Consistency: Maintain consistency in recording and tracking calibration intervals across all equipment.
- Proactive Management: Use a proactive approach to manage calibration schedules, ensuring that no equipment goes past its due calibration date.
- Integrated Systems: Leverage integrated calibration management systems to automate tracking and notification processes.
The procedure must include documentation of the calibration measurements prior to adjustment and measurements after any adjustments during calibration.
To ensure comprehensive control of testing, measuring, monitoring, and detection equipment, an organization must document calibration measurements both before and after any adjustments. This process enhances traceability and ensures the accuracy and reliability of equipment. Here’s how an organization can include this documentation in their calibration control process:
Steps to Include Documentation of Calibration Measurements
- Establish Clear Procedures:
- Calibration SOPs: Develop detailed standard operating procedures (SOPs) that specify how calibration measurements should be recorded. These SOPs should outline steps for both pre-adjustment and post-adjustment measurements.
- Documentation Templates: Create standardized templates or forms to capture all necessary information during calibration.
- Pre-Adjustment Measurements:
- Initial Reading: Before making any adjustments, record the initial measurements of the equipment. This step provides a baseline to understand how far the equipment has deviated from the standard.
- Documentation: Ensure that these measurements are documented accurately in the calibration log or record form.
- Adjustments:
- Record Adjustments: If adjustments are needed to bring the equipment into compliance, document the nature of these adjustments. This includes detailing what changes were made and why.
- Procedure for Adjustments: Include a step-by-step guide in the SOPs for making adjustments, ensuring consistency across all calibrations.
- Post-Adjustment Measurements:
- Final Reading: After adjustments, record the final measurements to confirm that the equipment now meets the required standards.
- Documentation: These post-adjustment measurements should also be documented clearly and compared with the pre-adjustment measurements.
- Verification and Approval:
- Review Process: Implement a review process where a qualified individual checks the calibration records to ensure accuracy and completeness.
- Approval: Have the calibration records approved by a supervisor or quality manager to validate the calibration process.
Integration into Quality Management System
- Quality Manual and SOPs:
- Include Calibration Documentation Requirements: Ensure that the quality manual and SOPs explicitly state the need for documenting both pre-adjustment and post-adjustment measurements.
- Update Regularly: Regularly update these documents to incorporate any changes in calibration processes or equipment.
- Training and Competency:
- Personnel Training: Train personnel on the importance of documenting calibration measurements accurately. Ensure they are familiar with the SOPs and templates.
- Competency Checks: Conduct periodic competency checks to ensure personnel follow the documented procedures correctly.
- Audit and Review:
- Internal Audits: Perform regular internal audits to verify that calibration records are maintained accurately and consistently.
- Continuous Improvement: Use audit findings to improve calibration processes and documentation practices.
Example of Calibration Record Form
| Equipment Type | ID Number | Calibration Date | Calibrated By | Pre-Adjustment Measurement | Adjustment Details | Post-Adjustment Measurement | Comments | Next Calibration Due |
|---|---|---|---|---|---|---|---|---|
| Pressure Gauge | PG-101 | 2024-01-10 | John Smith | 101 psi | Adjusted needle | 100 psi | N/A | 2025-01-10 |
| Temperature Sensor | TS-202 | 2023-12-15 | Jane Doe | 26.5°C | Recalibrated sensor | 25.0°C | N/A | 2024-06-15 |
| Multimeter | MM-303 | 2024-05-01 | John Smith | 12.2V | Adjusted dial | 12.0V | N/A | 2024-11-01 |
| Flow Meter | FM-404 | 2023-11-01 | Jane Doe | 102 L/min | Calibrated valve | 100 L/min | N/A | 2024-11-01 |
| Gas Detector | GD-505 | 2024-03-20 | John Smith | 0.52% | Replaced sensor | 0.50% | N/A | 2024-09-20 |
Key Points for Implementation
- Consistency and Accuracy: Ensure that all calibration measurements, both pre-adjustment and post-adjustment, are consistently and accurately documented.
- Traceability: Maintain clear traceability by linking each calibration record to the specific equipment through unique identification numbers.
- Clear Documentation: Use standardized forms and templates to ensure uniform documentation across all calibration activities.
- Regular Audits: Conduct regular audits to verify that the documentation process is being followed correctly and that records are complete and accurate.
- Continuous Training: Provide ongoing training to personnel to emphasize the importance of thorough documentation and adherence to procedures.
Calibration measurements prior to adjustment can be referred to as ‘as-found’. Calibration measurements after any adjustments can be referred to as ‘as-left’. When no adjustments are made, ‘as-found’ and ‘as-left’ are the same.
Calibration measurements taken prior to any adjustments are typically referred to as “as-found” measurements, while those taken after adjustments are referred to as “as-left” measurements. If no adjustments are made, the “as-found” and “as-left” measurements will be the same. This terminology helps clearly document the state of the equipment before and after calibration and provides a transparent record of the calibration process. Here’s how an organization can integrate this into their control of testing, measuring, monitoring, and detection equipment.
Integrating ‘As-Found’ and ‘As-Left’ Measurements
- Establish Clear Procedures:
- SOPs: Develop Standard Operating Procedures (SOPs) that clearly outline the requirement to record “as-found” and “as-left” measurements during the calibration process.
- Documentation Templates: Create templates for calibration records that include specific fields for both “as-found” and “as-left” measurements.
- Recording ‘As-Found’ Measurements:
- Initial Measurement: Take and document the initial measurement of the equipment before any adjustments are made. This “as-found” measurement serves as the baseline.
- Documentation: Use the designated section in the calibration record template to enter the “as-found” measurement.
- Adjustments:
- Document Adjustments: If adjustments are necessary, record the nature and extent of these adjustments. This documentation should detail what changes were made to correct any deviations.
- Procedure for Adjustments: Ensure that the SOPs provide step-by-step instructions on how to perform and document adjustments.
- 4. Recording ‘As-Left’ Measurements:
- Final Measurement: After adjustments, take a final measurement to confirm the equipment meets the required standards. This is the “as-left” measurement.
- Documentation: Enter the “as-left” measurement in the designated section of the calibration record template.
- No Adjustments Scenario:
- Same Values: If no adjustments are necessary, the “as-found” and “as-left” measurements will be the same. Document both as identical values.
Example of Calibration Record Form
| Equipment Type | ID Number | Calibration Date | Calibrated By | As-Found Measurement | Adjustment Details | As-Left Measurement | Comments | Next Calibration Due |
|---|---|---|---|---|---|---|---|---|
| Pressure Gauge | PG-101 | 2024-01-10 | John Smith | 101 psi | Adjusted needle | 100 psi | N/A | 2025-01-10 |
| Temperature Sensor | TS-202 | 2023-12-15 | Jane Doe | 26.5°C | Recalibrated sensor | 25.0°C | N/A | 2024-06-15 |
| Multimeter | MM-303 | 2024-05-01 | John Smith | 12.2V | Adjusted dial | 12.0V | N/A | 2024-11-01 |
| Flow Meter | FM-404 | 2023-11-01 | Jane Doe | 102 L/min | Calibrated valve | 100 L/min | N/A | 2024-11-01 |
| Gas Detector | GD-505 | 2024-03-20 | John Smith | 0.52% | Replaced sensor | 0.50% | N/A | 2024-09-20 |
Key Points for Implementation
- Consistency and Accuracy:
- Ensure that all “as-found” and “as-left” measurements are consistently and accurately documented.
- Use standardized forms to maintain uniformity in documentation across all calibration activities.
- Traceability:
- Maintain clear traceability by linking each calibration record to the specific equipment through unique identification numbers.
- Ensure that calibration records include the date, the technician who performed the calibration, and any adjustments made.
- Clear Documentation:
- Include detailed fields for “as-found” and “as-left” measurements in the calibration records.
- Document any adjustments made during calibration comprehensively.
- Regular Audits:
- Conduct regular audits to verify that the documentation process is being followed correctly and that records are complete and accurate.
- Use audit findings to improve calibration processes and documentation practices.
- Continuous Training:
- Provide ongoing training to personnel on the importance of thorough documentation and adherence to procedures.
- Emphasize the need to accurately record both “as-found” and “as-left” measurements.
By following these steps, an organization can effectively document “as-found” and “as-left” measurements, ensuring comprehensive control over testing, measuring, monitoring, and detection equipment. This practice supports compliance with API Q1 requirements and helps maintain high standards of equipment performance and product quality.
The procedure must include actions taken to prevent unintended use of TMMDE identified as out-of-calibration, beyond calibration interval, or not in-service.
To ensure the proper handling of Testing, Measuring, Monitoring, and Detection Equipment (TMMDE) identified as out-of-calibration, beyond calibration interval, or not in-service, here are the steps the organization can take:
Steps to Prevent Unintended Use of Out-of-Calibration TMMDE
- Identification and Segregation
- Tagging and Labeling: Clearly mark any TMMDE that is out-of-calibration, beyond its calibration interval, or not in-service with a prominent tag or label indicating its status.
- Segregation: Physically separate such equipment from calibrated and in-service TMMDE. Designate specific areas or storage locations for out-of-calibration and non-serviceable equipment to prevent accidental use.
- Documentation and Record-Keeping
- Log Records: Maintain a log that records the status of all TMMDE, including those out-of-calibration or not in-service. This log should include details such as the date the equipment was identified as out-of-calibration, the reason, and any actions taken.
- Update Records: Ensure that all changes in equipment status are promptly recorded in the inventory management system.
- Notification and Communication
- Internal Notification: Notify relevant personnel, including operators, technicians, and supervisors, about the status of the out-of-calibration TMMDE. Use internal communication channels such as emails, bulletin boards, or digital dashboards.
- Signage: Place clear signage in areas where TMMDE is used, reminding personnel to check the calibration status before using any equipment.
- Calibration and Maintenance Scheduling
- Scheduled Calibration: Develop and maintain a calibration schedule that ensures all TMMDE is calibrated at regular intervals. Track upcoming calibration dates to prevent equipment from going beyond its calibration interval.
- Preventive Maintenance: Include TMMDE in the preventive maintenance program to ensure regular checks and timely calibration.
- Access Control
- Restricted Access: Implement access controls to areas where TMMDE is stored. Limit access to authorized personnel who are aware of the calibration status and usage protocols.
- Sign-Out Procedures: Establish sign-out procedures for TMMDE, ensuring that equipment is checked for calibration status before it is used.
- Training and Awareness
- Staff Training: Regularly train staff on the importance of using calibrated equipment and the procedures for handling out-of-calibration TMMDE. Emphasize the potential risks and consequences of using such equipment.
- Awareness Programs: Conduct awareness programs to keep staff informed about calibration practices and the identification of out-of-calibration equipment.
Example of a Process for Handling Out-of-Calibration TMMDE
- Identification Process
- Routine Checks: Perform routine checks to identify any TMMDE that is out-of-calibration, beyond its calibration interval, or not in-service.
- Tagging: Attach a red “Out-of-Calibration” tag to any TMMDE identified as such.
- Segregation and Storage
- Dedicated Area: Move the tagged TMMDE to a designated quarantine area specifically for out-of-service equipment.
- Labeling: Ensure the storage area is clearly labeled and only accessible to authorized personnel.
- Documentation and Record-Keeping
- Log Entry: Make an entry in the TMMDE status log, recording the equipment ID, date identified, reason for out-of-calibration, and any immediate actions taken.
- Update Inventory: Update the equipment inventory to reflect the status change.
- Notification and Communication
- Internal Notification: Send a notification to relevant departments (e.g., quality control, production, maintenance) informing them of the status change.
- Post Signage: Place signs near workstations and storage areas, reminding staff to verify calibration status before use.
- Scheduling and Preventive Maintenance
- Calibration Scheduling: Schedule the out-of-calibration TMMDE for recalibration or maintenance as soon as possible.
- Preventive Checks: Include the equipment in the preventive maintenance schedule to ensure regular status checks.
- Access Control and Usage
- Restricted Access: Limit access to the quarantine area to maintenance and calibration personnel.
- Sign-Out Procedure: Implement a sign-out procedure that includes a calibration status check before any TMMDE is used.
- Training and Awareness
- Regular Training: Conduct training sessions for all relevant personnel on handling out-of-calibration TMMDE.
- Refreshers and Updates: Provide regular updates and refresher courses on the importance of using calibrated equipment.
Example of an Out-of-Calibration Log (Tabular Form)
| Equipment ID | Type | Date Identified | Status | Reason | Action Taken | Next Calibration Date | Responsible Person | Comments |
|---|---|---|---|---|---|---|---|---|
| PG-101 | Pressure Gauge | 2024-05-10 | Out-of-Calibration | Deviated readings | Tagged and moved to quarantine | Scheduled for 2024-06-01 | John Smith | N/A |
| TS-202 | Temperature Sensor | 2024-05-12 | Beyond Calibration Interval | Missed calibration schedule | Tagged and removed from service | Scheduled for 2024-05-20 | Jane Doe | N/A |
| MM-303 | Multimeter | 2024-05-15 | Not in-Service | Not required for current project | Stored in quarantine area | N/A | John Smith | Recalibration needed before next use |
| FM-404 | Flow Meter | 2024-05-17 | Out-of-Calibration | Inconsistent flow readings | Tagged and segregated | Scheduled for 2024-05-25 | Jane Doe | N/A |
| GD-505 | Gas Detector | 2024-05-20 | Not in-Service | Equipment retired | Documented and archived | N/A | John Smith | New equipment ordered |
By systematically tagging, segregating, documenting, communicating, scheduling, controlling access, and training personnel, an organization can effectively prevent the unintended use of out-of-calibration, beyond calibration interval, or not in-service TMMDE. This approach ensures compliance with API Q1 requirements and maintains the integrity of the product realization process.
The procedure must include when the TMMDE is found to be out of calibration, an assessment of the validity of previous measurements and actions to be taken on the TMMDE and product, including maintaining records and evidence of notification to the customer if suspect product has been shipped.
When equipment is found to be out of calibration, it is crucial for the organization to assess the validity of previous measurements and take appropriate actions to address any potential impact on the equipment and products. Here’s how the organization can include this assessment and related actions in their process, along with maintaining records and evidence of notification to the customer if suspect product has been shipped:
- Assessment of Validity of Previous Measurements
- Review Calibration Records: Examine the last known valid calibration date and compare it with the dates of the measurements taken using the equipment.
- Historical Data Analysis: Analyze historical measurement data to identify any anomalies or deviations that could indicate the time frame when the equipment may have started to produce inaccurate measurements.
- Impact Analysis: Assess the impact of potentially invalid measurements on the product’s quality and conformity. This involves determining which products were tested or measured using the out-of-calibration equipment and evaluating the risk associated with those products.
- Actions to Be Taken on the Equipment
- Immediate Segregation: Remove the out-of-calibration equipment from service immediately to prevent further use.
- Recalibration: Schedule the equipment for recalibration as soon as possible.
- Post-Recalibration Verification: After recalibration, verify the accuracy of the equipment before putting it back into service.
- Actions to Be Taken on the Product
- Product Recall or Hold: If there is a risk that suspect products were shipped, consider initiating a product recall or placing the products on hold until further investigation is complete.
- Retesting: Where feasible, retest or remeasure products that were tested with the out-of-calibration equipment using properly calibrated equipment.
- Quarantine: Quarantine any products that may have been affected by the out-of-calibration equipment until a thorough assessment is completed.
- Maintaining Records
- Incident Report: Document the incident, including the identification of the out-of-calibration equipment, the date it was found out-of-calibration, and the last valid calibration date.
- Assessment Records: Maintain detailed records of the validity assessment, including the analysis of historical data and the impact on product quality.
- Corrective Actions: Document all corrective actions taken, including recalibration, product recall, retesting, and any changes to processes to prevent recurrence.
- Notification to the Customer
- Customer Notification: If suspect products have been shipped, notify the customer immediately. Provide them with details about the potential issue, the risk involved, and the actions being taken to address it.
- Evidence of Notification: Maintain records of all communications with customers, including emails, letters, and phone call summaries, to provide evidence of the notification.
Example of a Process for Addressing Out-of-Calibration Equipment
- Detection and Initial Response
- Detection: Identify equipment that is out of calibration through routine checks or during use.
- Tag and Segregate: Tag the equipment as out-of-calibration and segregate it from other equipment.
- Assessment of Impact
- Review Records: Review calibration records and historical measurement data.
- Determine Impact: Identify products tested with the out-of-calibration equipment and assess the potential impact on product quality.
- Actions on Equipment
- Recalibration: Schedule the out-of-calibration equipment for immediate recalibration.
- Verification: Verify equipment accuracy post-recalibration before returning it to service.
- Actions on Products
- Product Recall: If necessary, initiate a product recall or hold products until further testing.
- Retesting: Retest products using calibrated equipment.
- Quarantine: Quarantine affected products until assessment is complete.
- Documentation and Communication
- Incident Report: Document the incident and the results of the impact assessment.
- Corrective Actions: Record all corrective actions taken.
- Customer Notification: Notify customers about the potential issue and actions taken. Maintain records of all communications.
Example of Records to Maintain
Incident Report (Tabular Form)
| Equipment ID | Type | Date Detected | Detected By | Last Calibration Date | Impact Assessment Summary | Actions Taken | Next Steps |
|---|---|---|---|---|---|---|---|
| PG-101 | Pressure Gauge | 2024-05-10 | John Smith | 2023-11-10 | Potential impact on batch 12345 | Equipment tagged and recalibrated | Retesting of batch 12345 |
| TS-202 | Temperature Sensor | 2024-05-12 | Jane Doe | 2023-12-12 | Reviewed 2 months of production data | Equipment removed from service | Notification to customers |
Customer Notification Log (Tabular Form)
| Customer ID | Customer Name | Product ID | Issue Description | Notification Date | Notified By | Customer Response | Follow-Up Actions |
|---|---|---|---|---|---|---|---|
| C-1001 | ABC Oil Co. | P-12345 | Potential pressure gauge issue | 2024-05-11 | John Smith | Acknowledged, awaiting further details | Recalibration and retesting results to be sent |
| C-1002 | XYZ Energy | P-67890 | Temperature sensor out of calibration | 2024-05-13 | Jane Doe | Requested immediate status update | Detailed impact assessment report to be provided |
By implementing a comprehensive process that includes assessment of the validity of previous measurements, actions on equipment and products, maintaining detailed records, and notifying customers, the organization can effectively manage the risks associated with out-of-calibration equipment. This ensures compliance with API Q1 requirements and maintains the integrity of the product realization process.
The procedure must include use of third-party, proprietary, employee-owned, and customer-owned TMMDE.
To ensure comprehensive control over all types of equipment used in product realization, an organization must address the use of third-party, proprietary, employee-owned, and customer-owned equipment within its documented procedures. Here’s how the organization can include these equipment types:
- Third-Party Equipment
- Identification: Clearly identify and record all third-party equipment used within the organization.
- Calibration and Certification: Ensure third-party equipment is calibrated and certified according to the organization’s standards and maintain records of calibration certificates.
- Usage Agreement: Establish agreements with third-party providers that outline the requirements for equipment maintenance, calibration, and performance standards.
- Verification: Verify the accuracy and suitability of third-party equipment before use and periodically thereafter.
- Proprietary Equipment
- Documentation: Maintain detailed documentation of proprietary equipment specifications, calibration requirements, and maintenance procedures.
- Calibration: Establish a calibration schedule that aligns with the equipment manufacturer’s guidelines and industry standards.
- Security: Implement measures to protect proprietary information associated with the equipment.
- Training: Ensure personnel are adequately trained in the use and maintenance of proprietary equipment.
- Employee-Owned Equipment
- Approval Process: Develop a formal approval process for the use of employee-owned equipment, including an initial inspection and verification of suitability.
- Calibration and Maintenance: Require employee-owned equipment to meet the same calibration and maintenance standards as organization-owned equipment. Maintain records of these activities.
- Agreement: Have employees sign an agreement that outlines their responsibilities for maintaining and calibrating their equipment according to organizational standards.
- Periodic Checks: Conduct periodic checks to ensure ongoing compliance with the organization’s standards.
- Customer-Owned Equipment
- Inventory Management: Keep a detailed inventory of all customer-owned equipment used within the organization.
- Inspection and Verification: Conduct initial and periodic inspections to verify the condition and calibration of customer-owned equipment.
- Documentation: Maintain records of customer-owned equipment, including calibration certificates, maintenance logs, and usage history.
- Communication: Establish clear communication channels with customers regarding the status and maintenance requirements of their equipment.
- Usage Authorization: Ensure that the use of customer-owned equipment is authorized and documented, with clear guidelines on handling and maintenance responsibilities.
The procedure must include maintenance.
To include the maintenance of equipment in the control of equipment, the organization must establish a detailed and documented procedure that encompasses all aspects of maintenance. This ensures equipment reliability, accuracy, and compliance with API Q1 requirements. Here are the key elements to include in such a procedure:
Key Elements to Include in the Maintenance of Equipment Procedure
- Identification and Documentation
- Equipment Inventory: Maintain a detailed inventory of all equipment requiring maintenance, including third-party, proprietary, employee-owned, and customer-owned equipment.
- Unique Identification: Assign a unique identification number or code to each piece of equipment for easy tracking and reference.
- Maintenance Records: Keep comprehensive records for each piece of equipment, documenting maintenance activities, schedules, and histories.
- Maintenance Schedules
- Frequency: Establish maintenance schedules based on manufacturer recommendations, industry standards, usage history, and risk assessments.
- Preventive Maintenance: Implement a preventive maintenance program to minimize equipment failures and extend equipment life. This should include routine inspections, cleaning, lubrication, and part replacements.
- Predictive Maintenance: Use predictive maintenance techniques, such as vibration analysis and thermal imaging, to identify potential issues before they lead to equipment failure.
- Maintenance Procedures
- Standard Operating Procedures (SOPs): Develop SOPs for each type of maintenance activity. These should detail the steps required to perform maintenance, the tools and materials needed, and safety precautions.
- Corrective Maintenance: Define procedures for addressing equipment failures or malfunctions, including troubleshooting, repair, and verification of functionality after repairs.
- Calibration Maintenance: Include specific procedures for maintaining the calibration of measuring and testing equipment, ensuring it remains within specified tolerances.
- Responsibilities and Training
- Personnel Responsibilities: Clearly define the roles and responsibilities of personnel involved in equipment maintenance, including technicians, supervisors, and managers.
- Training Programs: Provide training programs for maintenance personnel to ensure they have the necessary skills and knowledge to perform maintenance tasks effectively and safely.
- Monitoring and Reporting
- Condition Monitoring: Implement condition monitoring techniques to track the health and performance of critical equipment. Use data collected to inform maintenance decisions.
- Maintenance Logs: Maintain detailed logs of all maintenance activities, including the date, description of work performed, parts replaced, and the name of the technician who performed the work.
- Reporting: Establish reporting mechanisms for maintenance activities, including regular maintenance reports and summaries for management review.
- Spare Parts Management
- Inventory Management: Maintain an inventory of critical spare parts to ensure timely availability for maintenance and repairs.
- Part Replacement Records: Document all spare parts used during maintenance activities, including part numbers, quantities, and the equipment they were used on.
- Review and Improvement
- Periodic Review: Conduct periodic reviews of the maintenance program to identify areas for improvement and to ensure compliance with standards and regulations.
- Continuous Improvement: Implement a continuous improvement process to enhance maintenance practices based on feedback, performance data, and technological advancements.
Example of a Maintenance Record Log (Tabular Form)
| Equipment ID | Equipment Type | Last Maintenance Date | Next Scheduled Maintenance | Maintenance Activity | Technician | Parts Replaced | Comments |
|---|---|---|---|---|---|---|---|
| EQ-001 | Pressure Gauge | 2024-05-01 | 2024-11-01 | Calibration and cleaning | aa | None | Calibration within tolerance |
| EQ-002 | Temperature Sensor | 2024-04-15 | 2024-10-15 | Sensor inspection and testing | ss | Sensor probe | Replaced sensor probe |
| EQ-003 | Flow Meter | 2024-03-20 | 2024-09-20 | Flow calibration | zz | None | Calibration needed adjustments |
| EQ-004 | Employee-owned Laptop | 2024-05-05 | 2024-11-05 | Software update and testing | None | Updated to latest version | |
| EQ-005 | Customer-owned Pump | 2024-04-30 | 2024-10-30 | Performance testing and lube | hhh | Lubricant | Lubricated all moving parts |
Including the maintenance of equipment in the control of equipment involves establishing a comprehensive and documented procedure that covers identification, scheduling, detailed maintenance activities, responsibilities, monitoring, and continuous improvement. By doing so, the organization ensures that all equipment is maintained effectively, enhancing reliability and compliance with API Q1 requirements. This approach not only helps in preventing equipment failures but also supports the consistent delivery of quality products.
The procedure must include suitability for the planned monitoring and measurement activities.
To include suitability for the planned monitoring and measurement activities in the control of equipment, the organization must ensure that the equipment used is appropriate for its intended purpose and capable of providing accurate and reliable results. This involves several key steps:
Key Steps to Ensure Suitability
- Assessment of Requirements
- Identify Measurement Needs: Clearly define the monitoring and measurement requirements based on the product specifications, process requirements, and quality standards.
- Select Appropriate Equipment: Choose equipment that meets the accuracy, range, and resolution requirements for the intended monitoring and measurement activities.
- Qualification of Equipment
- Initial Verification: Perform an initial verification to ensure that the selected equipment meets the required specifications and is suitable for its intended use.
- Calibration: Ensure all equipment is calibrated to national or international standards before use. This includes establishing traceability to recognized standards.
- Validation: Validate the equipment’s performance under actual operating conditions to confirm it meets the necessary requirements.
- Documentation and Record Keeping
- Equipment Specifications: Maintain detailed records of equipment specifications, including accuracy, range, resolution, and any other relevant parameters.
- Calibration Certificates: Keep calibration certificates and records for all equipment, documenting calibration dates, results, and next calibration due dates.
- Validation Reports: Document the results of validation activities, including any tests conducted to verify the suitability of the equipment.
- Ongoing Monitoring and Maintenance
- Periodic Calibration: Establish a schedule for regular calibration of equipment to ensure ongoing accuracy and reliability.
- Performance Monitoring: Implement procedures to continuously monitor the performance of equipment, identifying any deviations or issues promptly.
- Maintenance Procedures: Develop and follow maintenance procedures to keep equipment in optimal condition, addressing wear and tear, and other potential issues.
- Training and Competence
- Operator Training: Ensure personnel are trained in the proper use, calibration, and maintenance of the equipment.
- Competence Assessment: Regularly assess the competence of personnel handling the equipment to ensure they can accurately perform the required monitoring and measurement activities.
- Review and Improvement
- Regular Review: Conduct regular reviews of equipment performance and suitability, using feedback from monitoring activities to identify areas for improvement.
- Continuous Improvement: Implement continuous improvement processes to enhance the suitability and performance of the equipment based on review findings and technological advancements.
Example of an Equipment Suitability Record (Tabular Form)
| Equipment ID | Equipment Type | Measurement Range | Accuracy | Calibration Date | Next Calibration Date | Validation Status | Validation Date | Validated By | Comments |
|---|---|---|---|---|---|---|---|---|---|
| EQ-101 | Digital Caliper | 0-150 mm | ±0.02 mm | 2024-01-15 | 2024-07-15 | Suitable | 2024-01-16 | aa | Validated for precision |
| EQ-102 | Pressure Gauge | 0-10 bar | ±0.1 bar | 2024-02-10 | 2024-08-10 | Suitable | 2024-02-11 | zz | Accurate within range |
| EQ-103 | Temperature Sensor | -50 to 150°C | ±0.5°C | 2024-03-20 | 2024-09-20 | Suitable | 2024-03-21 | Validated for full range | |
| EQ-104 | Flow Meter | 0.1-10 L/min | ±0.01 L/min | 2024-04-05 | 2024-10-05 | Suitable | 2024-04-06 | cc | Meets all specs |
Summary
Ensuring the suitability of equipment for planned monitoring and measurement activities involves careful selection, qualification, ongoing calibration, and maintenance of equipment. By including these aspects in the documented procedures, the organization can ensure that all equipment used is appropriate for its intended purpose and capable of providing accurate and reliable results, thus meeting the API Q1 requirements. This comprehensive approach helps maintain the integrity of the measurement processes and ensures consistent product quality.